Curtis ENABLE 40 User Manual

Page 16

Advertising
background image

10

Curtis enAble

®

40-Series Manual,

Rev. B

1741

HANDCONTROL

1740 POWERBASE

1742

ATTENDANT

CONTROL

1740 POWERBASE

4

1

5

2

3

Pin 1 B+

Pin 2 B-

Pin 3 Inhibit (active to ground/B-)

Pin 4 TXD

Pin 5 RXD

2 — INSTALLATION & WIRING

CONNECTING THE CONTROL MODULES
The communications cable on the 1741 user handcontrol and on the 1742
attendant control each terminate in a plug that fits into the powerbase’s 6-pin
connector. The attendant control’s cable includes a Y-junction into which the
user handcontrol’s cable connects, as shown in Figure 7.

5-pin charger/programmer port

The user handcontrol has a 5-pin port that accommodates a battery charger
and a 1311 programmer (although not at the same time). The charger plug’s
three pins fit into sockets 1, 2, and 3. The programmer plug’s two pins fit into
sockets 4 and 5. Both plugs are designed to snap securely into the port.

Allowable current for a charger in this port is 8 amps continuous (12 amps peak).

The inhibit pin (pin 3) halts all chair travel during charging. Additionally,

it can prevent specific actuator movements.

Fig. 7

Connecting the

control modules to the
powerbase.

(view looking

into the port)

Advertising