Warnings and precautions

Wherever the

symbol appears on the NucleoCounter instrument, it indicates that the

manual must be consulted for precautions and warnings.

Power and cables

Use the shielded USB and printer cables supplied by ChemoMetec A/S to ensure that you
maintain the appropriate EMI classification for the intended environment.

The USB interface connector of the NucleoCounter must only be connected to SELV

circuits. External computing devices connected to the USB interface connector of the
NucleoCounter has to comply with the standards, UL 1950 and IEC/EN 60950.

Table 1. Description of the NucleoCounter USB interface connector

The USB interface connector of the NucleoCounter must only be connected to SELV

circuits. External computing devices connected to the USB interface connector of the
NucleoCounter has to comply with the standards, UL 1950 and IEC/EN 60950.

Use the shielded printer cable supplied by ChemoMetec A/S to ensure that you maintain
the appropriate EMI classification for the intended environment.

The printer interface connector of the NucleoCounter must only be connected to the

printer supplied by ChemoMetec A/S. External computing devices connected to the
printer interface connector of the NucleoCounter has to comply with the standards, UL
1950 and IEC/EN 60950.

1

In normal operation mode (refers to Pin no 4)

Pin no.

Name

Maximum

Voltage level

1

1 +5V

+5

VDC

2

D-

+3.5

VDC

3

D+

+3.5

VDC

4

DGND

0

VDC

Metal

enclosure

Shield

(connected

to DGND)

0 VDC