8 Technical specifications

35

8 Technical specifications

8.1

Technical specifications for NucleoCounter SCC-100

Specificity

Somatic cells stained with the DNA specific fluorescent dye, PI.

Sample types

Milk samples containing dispersed somatic cells.

Sample consumption

The sample is mixed with a lysis buffer prior to analysis. It is
recommended that at least 500 µl of sample are used for this
dilution in order to reduce error caused by sample handling.
Approximately 50 µl of the lyzate mixture is loaded into the
cassette

Analysis volume

Approximately 2 µl of the lyzate mixture is being analyzed in
the NucleoCounter SCC-100, corresponding to approximately 1
µl of milk.

Somatic Cell Count in a volume of milk sample

Measurement range

Total range:
Recommended range:

10,000 to 2,000,000 cells/ml

100,000 to 750,000cells/ml

Presentation of Result

The NucleoCounter SCC-100 presents the estimated SCC in the
milk sample when milk and Reagent C are mixed in equal
amounts.

Operation

Menu-controlled by means of keyboard and LCD display.

Analysis time

When pressing “Run” on the NucleoCounter SCC-100 the result
will be displayed within 30 seconds.

Physical data

Weight
Height
Width
Depth

3 kg
26 cm
38 cm
22 cm

Power

See Section 0 Power Supply

Input
Fuse

12VDC (11-13VDC)
UL Listed fuse:
600mA/250Vac (UL 248)

Power consumption
NucleoCounter

Peak
Ready mode
Standby

25 W
2.5 W
2 mW

Operation conditions

Maximum relative humidity 80 percent for temperatures up to
31°C decreasing linearly to 65 percent relative humidity at
maximum 35°C; minimum temperature 15°C.

USB

USB, version 1.1.
Note: Does not support USB Hubs.

Printer

Thermal printer with RS-232 interface and cable optionally
supplied by ChemoMetec A/S.