9 Equipment and Accessories

39

UL listed, type SVT, rated min. 60C, 18 AWG, 3 conductors. Provided with molded-on
grounding-type (NEMA 5-15P) attachment plug, rated min. 125 Vac, min. 2.5A. Opposite
end terminates in molded on, IEC320 style connector, rated min. 125 Vac, min. 2.5 A.

For Europe 250Vac

Cord type min. H05RR-F or min. H03VV-F or min. H03VVH2-F, rated min. 60C, 3x0.75
mm2. Provided with molded-on grounding-type attachment Plug, rated min. 250 Vac,
min. 2.5 A. Opposite end terminates in molded on, IEC320 style connector, rated min.
250 Vac, min. 2.5 A.

Detachable power supply cord set for the specified Class II power supply:

For US 125 Vac

UL listed, type SPT-2 or SVT, rated min. 60C, 18 AWG, 2 conductors. Provided with
molded-on un-grounding-type (NEMA 1-15P) attachment Plug, rated min. 125 Vac, min.
2.5 A. Opposite end terminates in molded on, IEC320 style connector, rated min. 125
Vac, min. 2.5 A.

For Europe 250Vac

Cord type min. H05RR-F or min. H03VV-F or min. H03VVH2-F, rated min. 60C,
2x0.75mm2. Provided with molded-on un-grounding-type attachment Plug, rated min.
250 Vac, min. 2.5 A. Opposite end terminates in molded on, IEC320 style connector,
rated min. 250 Vac, min. 2.5 A.

The Mains supply cord and plug of the external power supply shall comply with any

national regulations.

The user shall be made aware of that, if the NucleoCounter and the external power

supply is used in a manner not specified by the manufacturer, the protection provided
by the NucleoCounter and the external power supply may be impaired.

Contact ChemoMetec A/S for information on order of a specific Power supply cord listed
above.

9.4

Dry compressed air

Dry, compressed air (KENAIR Air Duster - CFC Free) is used for removal of dust from the
optical system of the NucleoCounter (refer to chapter 6).

The item can be obtained from Kenro Ltd, Greenbridge Road, Swindon SN3 3LH, UK. Tel:
+44 (0) 1793 615836, E-mail: [email protected]. Alternatively the item, or a similar
item, might be available through local dealers.

9.5 External

Printer

An external impact or thermo printer may optionally be connected to the NucleoCounter
for the printing of test results.