5 Operation of the NucleoCounter

25

and with the lid closed. The check determines the number of CCD-pixels that show
elevated signal levels. If the number of pixels is above a predefined limit then the
cassette insertion slit can be contaminated with an interfering particle, like a dust
particle. An error message is shown in the display and a cleaning of the insertion slit is
recommended (refer to chapter 6 Maintenance of NucleoCounter).

If a PC with SomaticView is connected to the instrument a contaminant is usually
observed as an object, which is stationary in all images.

To initiate a

Zero Count Check type in “F50 + ↵”. Then the following text will appear in

the display:

0 count check

↵ Continue

Esc Abort

Figure 18 Zero Count Check, step 1

Press “↵” to continue. Now the instrument prompts you to remove any cassette from
the insertion slit:

Remove any

cassette

↵ Continue

Esc Abort

Figure 19 Zero Count Check, step 2

Press “↵” to continue. Now the instrument prompts you to close the lid in order to avoid
false light from entering the insertion slit during the Zero Count Check:

Close lid


↵ Continue

Esc Abort

Figure 20 Zero Count Check, step 3

Press “↵” to continue. Then a “Wait” message appears in the display:

wait...


Figure 21 Zero Count Check, step 3A