5 Operation of the NucleoCounter

26

After a few seconds the instrument has performed the Zero Count Check. If no
contamination of the insertion slit can be detected an “OK” message appears in the
display (see the figure below). Press any key to return to “Ready mode”.

0 count OK


Press any key

Figure 22 The result of the Zero Count Check is OK

If a contamination of the insertion slit is detected an error message is shown:

0 count ERROR

Refer to manual
Press any key

Figure 23 The Zero Count Check did not pass the test

If the Zero Count Check returns an error, it is recommended to clean the insertion slit,
e.g. with compressed, dry air (refer to chapter 6 Maintenance of NucleoCounter and
Technical Note No. 004 “

How to clean the NucleoCounter”

). Perform the Zero Count

Check again. If the error appears even after several attempts to clean the insertion slit
with compressed air and if necessary the NucleoCounter

Clean Kit, contact ChemoMetec

A/S or the local distributor.