ChemoMetec SCC-100 User Manual

Page 14

Advertising
background image

2

Table 2 Description of the NucleoCounter Printer Output connector

Pin no

Name

Maximum Voltage level

2

1

3

2 Rx

±10 VDC

3 Tx ±10

VDC

Metal Enclosure

DGND

0 VDC

2

1

3

Metal
Enclosure

Figure 1 Printer Output connector seen from the cable entry side.

NucleoCounter is powered by an external 12VDC power supply. For safe use, please
follow the instructions for connecting the power supply.

The NucleoCounter shall only be used with one of the following external power

supplies:

• Power supply Class I, Proton Electronic Industrial Co. Ltd., model SPN-260-12,

input rated 100-240 Vac, 50-60Hz, 1.6 A. Output rated 12 Vdc, 4.3 A.

• Power Supply Class I, Powerbox Europe AB, model EBH 03 131, input rated 100-

240 Vac, 47-63 Hz, 0.8A. Output rated 11-13 Vdc, 2.7 A.

• Power supply Class II, Celetron USA Inc., model ZVC40LT12E, input rated 100-240

Vac, 50-60Hz, 1.2 A. Output rated 12 Vdc, 3.3 A.

The detachable power supply cord set and appliance inlet of the external power

supply are considered as the disconnecting device.

The mains supply cord and plug of the external power supply shall comply with any

national regulations.

2

In normal operation. Refers to DGND Metal Enclosure. Taken from the datasheet for

MAX202ECSE.

3

There is no signal connected to pin 1.