25

Ensure that there are no objects on the weighing
plate. Press the

PRINT key.

In the display there appears the value of the
adjustment weight.

Place the required adjustment weight in the centre of
the sample dish, then press the

PRINT key. If the

adjustment was successful, this will be followed by
automatic reset to the menu.
Take away adjustment weight.
The

ESC key may be used to cancel the adjustment.

Errors during adjustment or the use of an incorrect
adjusting weight will result in an error message.
Repeat adjustment.

Return to weighing mode:
Press the

ESC-key repeatedly until the query

SAVE?“ appears.

Confirm query by pressing the

PRINT-key or reject it

by pressing the

ESC-key.

Setup

!

Saveing...

Setup

29.01.08

We recommend:
Carry out all adjustments as close as possible to the maximum load of the balance.