12.1 Remote control instructions

Commands Function
R CR LF

Reset to manufacturer setting - reset

PC CR LF

Command to call up the values of the balance

S CR LF

Transfer of stable weighing values in standard weighing unit

SI CR LF

Transfer of instable weighing values in standard weighing unit

SU CR LF

Last stable measuring value

SU CR LF

Current measuring value

Z CR LF

Zeroing stable value

ZI CR LF

Zeroing instable value

T CR LF

Taring stable value

TI CR LF

Taring instable value

C0 CR LF

Cancel continuous data output

C1 CR LF

Start continuous data output

NB CR LF

Serial no. of balance

FS CR LF

Weighing range max.

RV CR LF

Software issue

PD CR LF

Date display

PD CR LF

Time display

PMCR LF

Operation mode

PS CR LF

Balance parameters are printed out

B CR LF

Tone

A0 CR LF

Auto zero off

A1 CR LF

Auto zero on