KERN & SOHN MLS 50-3C User Manual

Page 50

Advertising
background image

50

MLS_C-BA-e-1330

Define
Settings are available for selection from 0 to 10
mg in 0.1 mg steps and from 5 seconds to 175 in 5
second steps.

Press

key; the cursor (►) marks the adjustable

parameter
Press the

- button anew and the current setting will

flash
Press the arrow keys (

 ), to carry out the desired

setting
Confirm by pressing the

PRINT key

Define 2
Drying is shut off as soon as the weight loss within
60 s is smaller than the set percentage value
(adjustable from 0.001% to 9.999%).

Press

key; the cursor (►) marks the adjustable

parameter

Press the

- button anew and the current setting will

flash

Press the arrow keys (

 ), to carry out the desired

setting

Confirm by pressing the

PRINT key

Test
Shut off for change of weighing value < 1mg in
freely defined time interval

Manual
Shutoff drying process manual

Temporal
Drying is shut off when set time has passed,
adjustable from 1min – 9h 59 min

Press

 -

key; the cursor (►) marks the adjustable

parameter
Press the

- button anew and the current setting will

flash
Press the arrow keys (

 ), to carry out the desired

setting
Confirm by pressing the

PRINT key

%

60s

% 0.100%