MLS_C-BA-e-1330

Begin temperature calibration:

 Close heating hood and carry out

temperature calibration as described in the
following.

 On the temperature calibration set switch-

on the digital thermometer by the ON/OFF
button.

Press the arrow keys (

 ), to select “

06 Drying

test

Press the

-key

The display to enter the temperature value to be
tested will appear.
The current menu item blinks.
Select the desired temperature with the navigation
keys.

Press PRINT button to accept.

Enter permitted deviation with the navigation keys.
Press PRINT button to accept.

Enter serial number of the temperature calibration set
with the navigation keys. Press PRINT button to
accept

The instrument will be automatically heated to the set
temperature. The current temperature and the
elapsed time will be indicated in the display.
The procedure lasts 8 minutes.

The display for entering the temperature of the
external thermometer will be displayed. Enter
temperature value with the navigation keys.
Press PRINT button to accept

If an optional printer is connected, a record will be
produced by pressing PRINT button.