KERN & SOHN MLS 50-3C User Manual

Page 30

Advertising
background image

30

MLS_C-BA-e-1330

04 Temperature adjustment
The temperature is measured at three points and can be adjusted at these
temperature points.
Before you do this, allow the device to cool down for at least 3 hours after the last
heating phase.

For preparation/installation of temperature calibration set( MLB-A12) , see

function "06 Temperature calibration“

Begin temperature adjustment:
Select by using the arrow keys (

 ) „

04 Temp.cal

Press

 key

Temperature calibration starts.

Temperature calibration of first point will take 8 min
after which you will hear an acoustic signal.

Enter the temperatur value as indicated on the
external thermometer with the help of the navigation
buttons

 .

Confirm with

PRINT key, the second heating phase

is starting.

Temperature calibration of secend point will take 8
min after which you will hear an acoustic signal

Enter the temperatur value as indicated on the
external thermometer with the help of the navigation
buttons

 .

Confirm with

PRINT key, the third heating phase is

starting.

Setup

29.12.04

Setup

Setup

29.01.08

Setup

Temperature calibration
1 point 14:59

Setup

29.01.08

Setup

Temperature calibration
1 point 00:00

Setup

29.01.08

Setup

Temperature calibration
Set temp. value [°C]
25

Setup

29.01.08

Setup

Temperature calibration
Set temp. value [°C]
24

Setup

29.01.08

Setup

Temperature calibration
2 point 14:59

Setup

29.01.08

Setup

Temperature calibration
2 point 00:00

Setup

29.01.08

Setup

Temperature calibration
Set temp. value [°C]
120

Setup

29.01.08

Setup

Temperature calibration
Set temp. value [°C]
122

Setup

29.01.08

Setup

Temperature calibration
2 point 00:00

1 point 07:59

2 pointt 07:59

90

89

3 point 07:59