MLS_C-BA-e-1330

To switch the display to the various units of the
measured result, press the

M-key.

When drying is finished, you will hear an acoustic
signal and the heating will be shut off.

The display shows the measured result in the set
unit.

Open lid and remove the sample with the help of
the removal aid.
Caution: Caution! Sample dish and all parts of the
sample chamber are hot!

Data output via RS 232
interface is achieved by
pressing the

PRINT-key:

--------------- Start drying----------------

Program no. : 1

Name : Floar 100

Profile : Standard

Dry temp. : 120 °C

Switch off : Automatical 3

Result : Moisture contents -

%M

Printout int : 30 s

Target : None

T. coeficjent : 0.00

Start weight : 9.993 g

0:00:30 0.01 %M

0:01:00 0.02 %M

0:01:30 0.03 %M

Final mass : 9.990 g

Dry time : 0:01:45

Final result : 0.03 %M

--------End of drying procedure------