KERN & SOHN MLS 50-3C User Manual

Page 26

Advertising
background image

26

MLS_C-BA-e-1330

03 Calibration test

This is used to determine the deviation from the last adjustment. This is only a
check, i.e. no values are changed.

Press the arrow keys (

 ), to select “

03 Cal. test“.

Press

 button

Ensure that there are no objects on the weighing
plate.

Press

PRINT key, on the display appears the weight

value of the required adjusting weight.

Then place the required adjustment weight in the
centre of the sample dish, press the

PRINT -key.

The result of the calibration test is shown on the
display.

Take away adjustment weight. To return to the menu
or weighing mode, press the

ESC-key several times.

Return to weighing mode:
Press the

ESC-key repeatedly until the query

SAVE?“ appears. Confirm query by pressing the

PRINT-key or reject it by pressing the ESC-key.

Setup

29.12.04

Setup

Einstellungen

Calibration check

Cear the pan

[Enter]

Einstellungen

Calibration check

Weight measuring

Einstellungen

Calibration check

Load weigh

[Enter]

t 50.0g

Einstellungen

Calibration check

Weight measuring

Calibration

Check results

Cal.: 180.1354

Act.: 180.1438

Diff.: - 0.0084