14 instant help – KERN & SOHN MLS 50-3C User Manual

Page 82

Advertising
background image

82

MLS_C-BA-e-1330

14 Instant help


Fault

Possible cause

Display is not lit up.

• The display unit is not switched on
• The mains supply connection has been

interrupted (mains cable not plugged in/faulty)

• Power supply interrupted
• Fuse has blown

The display does not change
when a sample is being loaded

• Sample dish / dish holder is fitted incorrectly.

The weight display changes
constantly / the stability display
does not appear.

• Sample dish has contact with wind protection

device or heated cover

• Draught/air movement
• Table/floor vibrations
• Electromagnetic fields / static charging (choose

different location/switch off interfering device if
possible)

Incorrect measuring result

• Check adjustment
• No resetting to zero before loading the sample

Measurement is taking too
long

• Incorrect shut-down criteria set

Measurement is
not reproducible

• Sample is not homogenous
• Drying time is too short
• Drying temperature too high (e.g. oxidation

sample material, boiling point of sample
exceeded)

• Temperature sensor soiled or defective

Drying does not start

• Heated cover open
• The mains supply connection has been

interrupted (mains cable not plugged in/faulty)