MultiLab 4010-2

What to do if...

ba76142e03

08/2013

117

14.3 Dissolved oxygen

Error message,

OFL

Error message,

Error

14.4 Conductivity

Error message,

OFL

Error message,

Error

More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.

Cause

Remedy

– Measured value outside the measur-

ing range

– Select a suitable measuring

medium

Cause

Remedy

– Sensor contaminated

– Clean the sensor

– Measured temperature value outside

the operating conditions
(display of OFL/UFL instead of a
temperature value)

– Keep to the temperature range

for the test sample

– Defective sensor

– Calibrate
– Change sensor cap
– Replace sensor

More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.

Cause

Remedy

– Measured value outside the measur-

ing range

– Use suitable IDS conductivity

sensor

Cause

Remedy

– Sensor contaminated

– Clean the sensor and replace it

if necessary

– Calibration solution not suitable

– Check the calibration solutions