10.4 D.O. measurement settings

10.4.1 Settings for D.O. measurement

The settings are available in the menu for measurement and calibration set-
tings. To open the settings, display the required measured parameter in the
measured value display and press the <ENTER> key. After completing the set-
tings, switch to the measured value display with <M>.

10.4.2 Enter Sensor Cap coefficients

10.4.3 Saturation local

Irrespective of the height or air pressure, the calibration value is set to 100%.
Saturation local is ideal for EU compliance.

When Saturation local is enabled, an L will appear next to D.O.% on the
display.

D.O. mg/L readings are unaffected by the selection of Saturation local.

10.5 Cond measurement settings

10.5.1 Settings for IDS conductivity sensors

The settings are made in the menu for the measured parameter, conductivity.
To open the settings, display the required measured parameter in the mea-
sured value display and press the <ENTER> key. After completing the settings,
switch to the measured value display with <M>.
The possible settings are individually displayed for each sensor. Below the set-
ting menu is displayed for two IDS sensors (

4310, 4320).

Default settings are printed in bold.

Setting menu of

4310

The values of the coefficients are provided with the sensor cap.

1.

Change the digit of the highlighted position with <><>.

2.

Go to the next position with <><>.

3.

Confirm with<ENTER> when a coefficient is completely entered.

Menu item

Possible
setting

Explanation

Calibration /Calibra-
tion record

-

Displays the calibration record of the
last calibration.