pH value

MultiLab 4010-2

36

ba76142e03

08/2013

5.2.6 Calibration data

Displaying the cali-

bration data

The calibration record of the last calibration is to be found under the menu item,
Calibration / Calibration record. To open it in the measured value display, press
the <CAL_> key.

The calibration records of the last 10 calibrations are to be found in the menu,
Calibration / Calibration data storage / Display. To open the Calibration menu,
press the <ENTER> key in the measured value display.

Calibration

Determined val-
ues

Displayed calibration data

1-point

Asy

Zero point = Asy

Slope = Nernst slope
(-59.2 mV/pH at 25 °C)

2-point

Asy
Slp.

Zero point = Asy

Slope = Slp.

3-point to 5-
point

Asy
Slp.

Zero point = Asy

Slope = Slp.

The calibration line is calculated by
linear regression.

You can display the slope in the units, mV/pH or % (see section
10.1.1 S

ETTINGS

FOR

P

H

MEASUREMENTS

, page 83).

The calibration record is automatically transmitted to the interface
after calibrating.