2 buffer sets for calibration, See section 10.1.2 b, Uffer – YSI MultiLab 4010-2 User Manual

Page 85: Sets, Calibration

Advertising
background image

MultiLab 4010-2

Settings

ba76142e03

08/2013

85

10.1.2 Buffer sets for calibration

You can use the buffer sets quoted in the table for an automatic calibration. The
pH values are valid for the specified temperature values. The temperature
dependence of the pH values is taken into consideration during the calibration.

No.

Buffer set *

pH values

at

1

YSI *

4.000
7.000
10.000

25 °C

2

ConCal

Any

Any

3

NIST/DIN
DIN buffers according to DIN 19266
and NIST Traceable Buffers

1.679
4.006
6.865
9.180
12.454

25 °C

4

TEC
Technical buffers

2.000
4.010
7.000
10.011

25 °C

5

Merck 1*

4.000
7.000
9.000

20 °C

6

Merck 2 *

1.000
6.000
8.000
13.000

20 °C

7

Merck 3 *

4.660
6.880
9.220

20 °C

8

Merck 4 *

2.000
4.000
7.000
10.000

20 °C

9

Merck 5 *

4.010
7.000
10.000

25 °C

10

DIN 19267

1.090
4.650
6.790
9.230

25 °C

11

Mettler Toledo USA *

1.679
4.003
7.002
10.013

25 °C

Advertising