MultiLab 4010-2

pH value

ba76142e03

08/2013

35

Continuing with

three- to five-point

calibration

5.2.5 Calibration points

Calibration can be performed using one to five buffer solutions in any order
(single-point to five-point calibration). The meter determines the following
values and calculates the calibration line as follows:

20. Thoroughly rinse the sensor with deionized water.

21. Immerse the sensor in the next buffer solution.

22. When measuring without temperature sensor:

Enter the temperature of the buffer with <><>.

23. Start the measurement with <ENTER>.

The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.

24. Wait for the measurement with stability control to be completed or ter-

minate the stability control and take over the calibration value with
<ENTER>.
The pH value of the buffer solution is displayed.

25. Set the nominal buffer value for the measured temperature with

<><>.

26. Accept the calibration value with <ENTER>.

The calibration display for the next buffer appears (voltage display).

27. If necessary, use <M> to finish the calibration.

The calibration record is displayed.
or
Continue calibrating using the next buffer with <ENTER>.

After the fifth buffer has been measured the calibration is automat-
ically finished. Then the calibration record is displayed.

The calibration line is determined by linear regression.

03.04.2013 08:00