10 Settings

10.1 pH measurement settings

10.1.1 Settings for pH measurements

The settings are made in the menu for calibration and measurement settings of
the pH/ORP measurement. To open the settings, display the required mea-
sured parameter in the measured value display and press the <ENTER> key.
After completing the settings, switch to the measured value display with <M>.
Default settings are printed in bold.

Menu item

Possible
setting

Explanation

Calibration /Calibration
record

-

Displays the calibration record of the last calibration.

Calibration / Calibration
data storage
/ Display

-

Displays the last calibration records (max. 10)

Calibration /
Calibration data storage
/ Output to RS232/USB

-

Outputs the stored calibration data to the USB-B interface
(PC)

Calibration / Calibration
data storage
/ Output to
USB flash drive or
printer

-

Outputs the stored calibration data to the USB-A interface
(USB memory device/USB printer)

Calibration /Buffer

YSI
ConCal
NIST/DIN
...

Buffer sets to be used for pH calibration.
More buffers and details: see section 10.1.2 B

UFFER

SETS

FOR

CALIBRATION

, page 85 and section 5.2

P

H

CALIBRATION

,

page 29.

Calibration /Single-point
calibration

yes
no

Quick calibration with 1 buffer

Calibration /Calibration
interval

1 ... 7 ... 999
d

Calibration interval for the IDS pH sensor (in days).
The meter reminds you to calibrate regularly by the flash-
ing sensor symbol in the measuring screen.

Calibration /Unit for
slope

mV/pH
%

Unit of the slope.
The % display refers to the Nernst slope of
-59.2 mV/pH (100 x determined slope/Nernst slope).

QSC /First calibration

-

Starts the initial calibration with QSC buffers.
This menu item is only available as long as no initial cali-
bration was carried out with the connected IDS sensor.

QSC /Record of first cal-
ibration

-

Displays the calibration record of the QSC initial calibra-
tion.

QSC /Control calibra-
tion

-

Starts the control calibration with QSC buffers.
This menu item is only available if an initial calibration was
carried out with the connected IDS sensor.