6 Cleaning and Maintenance

English

11

Cleaning

Cautions

• Disconnect the LCSU 4 from external power prior to

cleaning.

• Use a minimum of liquid to prevent any electrical shock

hazard. Do not immerse the LCSU 4, or allow it to stand

in water or other liquids. This can damage the device and/

or cause electrical shocks resulting in injury to persons.

Warning

Do not pump any cleaning solution or other liquids

through the vacuum pump, i.e. through the Vacuum

Connector. This can damage the LCSU 4.

Main Cabinet
1

Disconnect from external power supply.

2

Clean Cabinet surfaces by carefully wiping with a soft cloth

or sponge with mild detergent. Use hand dishwashing liquid

or similar that is compatible with the Material Chart (see

Chapter 10).

3

Dry all surfaces using a clean cloth or paper towel.

Canisters and Patient Tubes
Dispose after use.

Important
The Canisters and Patient Tubes are all disposable items.

Do not attempt to clean or reuse these. Due to the risk

of cross contamination, all disposable items must be

replaced after each use. They are for single-patient use

only.

Vacuum Tube (for 800 ml version) and Wire Stand
Wash by immersing and scrubbing in liquid detergent and

water solution.
1

Rinse thoroughly in pure water.

2

Allow to dry. Disinfect if desired.

High Efficiency Filter Kit (for 800 ml Version)
• Filter cannot be cleaned or disinfected.
• Replace Filter immediately if contamination or

discoloration is observed, or if it gets wet.

* If the unit is used on patients in areas where cross

contamination is an issue, it is recommended that the Filter is

replaced after each use.

Carry Bags
Wipe Bags according to instructions provided above for Main

Cabinet. Do not launder.

Device Test

After each reassembly, and before returning a LCSU 4 to

operational use, the Device Test should be performed:

1

Start the test with a fully charged Battery

2

Turn the unit “ON”

3

Block the Patient Tube

4

Set Suction Vacuum level to 550 mmHg

5

Unblock and then Block the Patient Tube again.

6

Result: Unit should return to 550 mmHg setting.

7

Repeat procedure for 300 mmHg and 50 mmHg settings.

Cautions
Do not attempt to use any LCSU 4 that has not passed

the above test. If the LCSU 4 unit does not test

satisfactorily, recheck all parts of the assembly and test

once again. If necessary, contact Laerdal Medical or one

of its authorized distributor.

Disposal

When discarding the LCSU 4, we recommend it be discarded

according to local protocol.

This appliance is marked according to the European directive

2002/96/EC on Waste Electrical and Electronic Equipment

(WEEE). By ensuring this product is disposed of correctly,

you will help prevent potential negative consequences for the

environment and human health, which could otherwise be

caused by inappropriate waste handling of this product.

The symbol on the product, or on the documents

accompanying the product, indicates that this appliance

may not be treated as household waste. Instead it shall

be handed over to the applicable collection point for the

recycling of electrical and electronic equipment.

Disposal must be carried out in accordance with local

environment regulations for waste disposal.

For more detailed information about treatment, recovery

and recycling of this product, please contact your local city

office, your household waste disposal service or the Laerdal

representative where you purchased the product.