3 LCSU 4 Setup

English

6

Prepare the LCSU 4 for Operation

Important
All models are shipped with the Battery inside the unit, but not connected. Connect the Battery and charge it fully before

using the unit. See Charging Instructions (Chapter 5).

Assemble 800 ml Version

1

2

3

A

B

Canister Ports
A

- Vacuum Port

B

- Patient Port

1

Secure the Lid to the

Canister.

2

Place the Canister into

Wire Stand.

3

Ensure that the Patient

Port is accessible.

4

5

6

7

4

Attach the Connector

to the Vacuum Port

Connector on the Suction

Unit.

5

Attach the WHITE

Connector to the Vacuum

Port Connector on the

Canister.

6

Check that all Vacuum

Tube connections are

firmly attached.

7

Attach the Patient Tube

to Patient Port on the

Canister.

Important
The 800 ml Canister has an internal Filter in the Lid. The Canister is disposable and cannot be cleaned. The Filter

automatically stops suction/flow when the Canister is full, or the Filter becomes saturated if the unit tips onto its side during

use.

Caution
Always use the 800 ml Canister supplied by Laerdal, which has an internal Filter. Never connect any type of Patient Tubing

directly to the LCSU 4 Vacuum Inlet Port Connection. Overflow of suctioned material into the LCSU 4 pump will result in

loss of suction and permanent damage to the unit. In the event of overflow, do not use the LCSU 4. Contact Laerdal Medical

or your authorized distributor.

IN

OUT

Operation with High Efficiency Filter Kit

To increase filtration efficiency the Vacuum Tube may be replaced with a High Efficiency Filter Kit

(Cat. No. 886116).

Ensure Filter IN points towards the WHITE connector, and OUT towards the BLUE. For Cleaning

and Maintenance, see Chapter 6.