3 LCSU 4 Setup

English

7

Assemble 300 ml Version

1

Push the upper port connection into the Vacuum Inlet and

check that the lower part of the Canister clicks in place.

2

Connect the Patient Tube to the Patient Port. Ensure that

all connections are secure to prevent leakage.

1

2

300ml

250ml

200ml

150ml

100ml

50ml

300ml

250ml

200ml

150ml

100ml

50ml

Important

• The 300 ml Canister is sealed, with an internal Filter. The

Canister is disposable and cannot be cleaned. The Filter

automatically stops suction/flow when the Canister is full,

or the Filter becomes saturated if the unit tips onto its side

during use.

• The 300 ml Canister (Cat. No. 886100) can also be used with

the previous model LCSU 3. However, the LCSU 4 cannot

be used with LCSU 3 Canisters.

• To improve the upright stability, a Wire Stand (Cat. No

886115) is offered as an Optional Accessory. This adds

both a handle and a foot to the unit.

Caution
Do not attempt to install or use LCSU 3 Canisters on

LCSU 4.

Check Before Each Use

1

The Unit should not be damaged.

2

The Unit should be clean.

3

All parts should be properly assembled (Canister, Tubes etc.).

4

Perform Device Test after each reassembly (see Chapter 6).

5

Check Battery level: While performing the Device Test the

Battery level should not illuminate RED. If the Battery

level illuminates RED, the Battery should be charged. See

Charging Instructions (Chapter 5).

Important
For both models - always have an extra Canister available

in case the first Canister is filled completely, or the unit

becomes tipped on to its side and the Filter becomes

saturated and stops the suction/flow.

Warning

If the Canister on either model fills and the shut-off

mechanism activates, and you have no spare Canister

ready for immediate replacement, shut off LCSU 4 and

utilize alternative methods according to local protocol

to clear patient’s airway. Continued efforts to suction

with LCSU 4 Canister(s) filled may cause overflow that

will prevent suctioning, damage the pump, void the unit

warranty, and lead to prolonged downtime.

Cat. No. 886115