BINDER BD 23 User Manual

Page 25

Advertising
background image

BD / ED / FD (E2) 02/2015

page 25/63

6.2 Display / entry of temperature set-point (with selected temperature ramp)

If previously a temperature ramp value has been selected (chap. 6.4.2):

Press button

in normal display / actual value display during ramp operation to have displayed the

actual temperature ramp set-point, which changes according to the selected gradient, in addition to the

entered target set-point for temperature.
The chamber is operating, the controller is in normal display (actual value

display). The

actual temperature value (example: 22 °C) is displayed:

1. Press

button

The display shows alternately “

SPr” and the actual temperature ramp set-point changing according

to the selected gradient (example: 42 °C):

This ramp set-point is displayed only, it is not adjustable.

2. Press

button

The display shows alternately “

SP” and the previous temperature set-point (example: 60 °C):

3. With the

buttons enter a set-point value between 0 and 300.

The desired temperature set-point can be selected in a temperature range from 5 °C / 9 °F

above room temperature up to 100 °C / 212 °F (BD) or 300 °C / 572 °F (ED/FD).


4. Wait 2

seconds until the entered temperature value

is

taken over

(display flashing once).

5. Press

button to return to normal display / actual value display (automatically after approx. 30

seconds).

Adjust the safety device following any changes of the set-point (chap. 7).