4 user level settings – BINDER BD 23 User Manual

Page 29

Advertising
background image

BD / ED / FD (E2) 02/2015

page 29/63

6.4 User level settings

By pressing down button

in normal display (actual value display) for 5 sec, you enter the user menu.

Settings in this menu affect controller operation.

User level overview:

Press down button

for approx. 5 seconds

Select the

temperature unit (chap. 6.4.1)

Press

button

Ramp function (chap. 6.4.2)

Press

button

Setting the

unit address (chap. 6.4.3)

Figure 6: Press down

button for approx. 5

seconds

Press button

to return to normal display with display of the temperature set-point.

Or:

After approx. 30 seconds the controller automatically returns to normal display / actual value display.
All settings can be carried out independently (as described in he individuel sections) or one after the other

during one single process.

The defined parameters are not deleted when the main power switch is turned off or in case

of power failure.

6.4.1 Temperature unit change between degrees Celsius °C and degrees Fahrenheit °F

If required, the temperature display can be changed as follows:

1. Press down

button for approx. 5 seconds.

The display alternately shows “

unit” and the actual setting of the temperature unit:

2. Use the

buttons to set the required unit.

3. The set unit is automatically adopted after 2 seconds.

C = degrees Celsius
F= degrees Fahrenheit

0 °C = 31 °F
100 °C = 212 °F

Conversion:
[Value in °F] = [Value in °C] ∗ 1.8 + 32

When specifying the set point ramp (see chap. 6.4.2) this setting is accordingly taken as the basis.

If the unit is changed, the temperature set-point and limits are converted accordingly.