12. Technical description

12.1 Factory calibration and adjustment

This unit was calibrated and adjusted in the factory. Calibration and adjustment were performed using

standardized test instructions, according to the QM DIN EN ISO 9001 system applied by BINDER

(certified since December 1996 by TÜV CERT). All test equipment used is subject to the administration of

measurement and test equipment that is also constituent part of the BINDER QM DIN EN ISO 9001

systems. They are controlled and calibrated to a DKD-Standard at regular intervals.

12.2 Definition of usable volume

The usable volume illustrated below is calculated as follows:

b

c

c

a

b

a

C

B

A

A, B, C = Internal dimensions (W, H, D)

a, b, c = Wall clearances

a = 0.1 x A

b = 0.1x B

c = 0.1 x C

V

USE

= (A - 2a) x (B - 2b) x (C - 2c)

Figure 11: Determination of the useable volume

The technical data refers to the defined usable volume.

Do NOT place samples outside this usable volume.
Do NOT load this volume by more than half to enable sufficient airflow inside the chamber.
Do NOT divide the usable volume into separate parts with large area samples.
Do NOT place samples too close to each other in order to permit circulation between them

and thus obtain a homogenous distribution of temperature.

12.3 Over current protection

Single-phase devices are protected by a miniature fuse against over current, accessible from the

outside. The miniature fuse is located at the rear of the chamber below the strain relief of the power cord.

The fuse holder is equipped with a fuse clip 5mm x 20 mm. (CUL-Version 6,3x32 mm). The fuse may be

replaced only with a substitute of the same ratings. Refer to the technical data of the respective device

type.
Three-phase devices are equipped with internal fuses not accessible from outside. If these fuses are

blown, please inform an electronic engineer or BINDER service.