Recommended precautions: To protect the eyes use sealed protective goggles. Suitable protective

gloves with full contact: butyl or nitrile rubber, penetration time >480 minutes.

CAUTION

Contact with skin, ingestion.
Skin and eye damage due to chemical burns.

Do not ingest. Keep away from food and beverages.

Do NOT empty into drains.

Wear protective gloves and goggles.

Avoid skin contact.

9.2.2 Decontamination

The operator must ensure that proper decontamination is performed in case a contamination of the

chamber by hazardous substances has occurred.
Disconnect the chamber from the power supply prior to decontamination. Pull the power plug.
Do not use decontamination agents that may cause a hazard due to reaction with components of the

device or the charging material. If there is doubt regarding the suitability of cleaning products, please

contact BINDER service.
You can use the following disinfectants:

Inner chamber

Standard commercial surface disinfectants free from acid or halides.
Alcohol-based solutions.
We recommend using the disinfectant spray Art. No. 1002-0022.

For chemical disinfection, we recommend using the disinfectant spray Art. No. 1002-0022.
Any corrosive damage that may arise following use of other disinfectants is excluded from

liability by BINDER GmbH.

With every decontamination method, always use adequate personal safety controls.

In case of impurity of the interior with biological or chemical hazardous material, there are three possible

procedures depending on the type of contamination and of the charging material.
1. The drying and heating ovens ED and FD can be hot air sterilized at 190 °C / 374 °F for at least 30

minutes. All inflammable goods must be removed from the interior before. With the incubators BD it is

possible to perform a hot-air disinfection at 100 °C / 212 °F.

2. Spray the inner chamber with an appropriate disinfectant.

Before start-up, the unit must be absolute dry and ventilated, because explosive gases may form

during the decontamination process.

3. If necessary, have strongly contaminated inner chamber parts removed by an engineer for cleaning, or

have them exchanged. Sterilize the inner chamber parts in a sterilizer or autoclave.

In case of eye contact, the disinfectant spray may cause eye damage due to chemical

burns. Follow the operating instructions and safety hints labeled on the bottle of the

disinfectant spray.