Unit size

BD 23 BD 53 BD 115 BD 240 BD 400 BD 720

Electrical data
IP system of protection acc. to EN

60529

IP

20

20

20

20

20

20

Nominal voltage (±10 %) 50/60 Hz

V

230 1N~ 230 1N~ 230 1N~ 230 1N~ 230 1N~ 230 1N~

Nominal power

kW

0.2

0.4

0.4

0.68

0.85

1.25

Energy consumption at 37°C / 98.6°F Wh/h

11

11

20

33

56

80

Unit fuse 5 x 20 mm

230V / 10A / middle-time-lag (M)

Amp

10

external

10

external

10

external

10

external

10

external

10

external

Power plug

shock proof plug

Installation category acc. to IEC 61010-1

II

II

II

II

II

II

Pollution degree acc. to IEC 61010-1

2

2

2

2

2

2

Electrical connection data BD-UL acc. to cUL standard (for USA and Canada):

Unit size

BD 23-

UL

BD 53-

UL

BD 115-

UL

BD 240-

UL

BD 400-

UL

BD 720-

UL

Electrical data
Nominal voltage (±10%) 60 Hz

V

115 1N~ 115 1N~ 115 1N~ 115 1N~ 115 1N~ 115 1N~

Nominal power

kW

0.20

0.30

0.40

0.68

0.85

1.25

Nominal current

Amp

1.8

2.6

3.5

5.9

7.4

10.9

Power plug

NEMA 5-15P 5-15P 5-15P 5-15P 5-15P 5-20P

Unit fuse 6,3 x 32 mm 250V / super-

time-lag TT

Amp

12,5

external

12,5

external

12,5

external

12,5

external

12,5

external

16

external

Additional temperature fuse class 1

(DIN 12880)

internal internal internal internal internal internal

Legend: 1) without outer glass door 2) up to 98 % of the set value
All technical data is specified for unloaded units with standard equipment at an ambient temperature of

+25 °C / 77 °F and a power supply voltage fluctuation of ±10. The temperature data is determined in

accordance to BINDER factory standard following DIN 12880, observing the recommended wall

clearances of 10 % of the height, width and depth of the inner chamber.
All indications are average values, typical for units produced in series. We reserve the right to

change technical specifications at any time.

If the cabinet is fully loaded, the specified heating up times may vary according to the load.

With option interior socket: If electrical devices are connected and operating inside the

chamber, the temperature range may be modified due to heat emission.

12.5 ED technical data

Unit size

ED 23 ED 53 ED 115 ED 240 ED 400 ED 720

Exterior dimensions
Width

mm

inch

435

17.13

635

25.00

835

32.87

1035

40.75

1235

48.62

1235

48.62

Height (incl. feet/castors)

mm

inch

495

19.49

620

24.41

705

27.76

825

32.48

1025

40.35

1530

60.24

Depth

mm

inch

520

20.47

575

22.64

645

25.39

745

29.33

765

30.12

865

34.06

incl. door handle and exhaust duct

mm

inch

105

4.13

105

4.13

105

4.13

105

4.13

105

4.13

105

4.13

Wall clearance rear

mm

inch

100

3.94

100

3.94

100

3.94

100

3.94

100

3.94

100

3.94