Function:
The temperature safety device is functionally and electrically independent of the temperature control

system and if an error occurs it performs a regulatory function.
If you turn the control knob (7) to its end-stop (position 10), the safety device class 3.1 protects the

chamber. If you set it to a temperature a little above the controller’s set-point temperature, it protects the

charging material.
If the safety device has taken over control (identifiable by the red alarm lamp (7a) lighting up and, in case

of the option audible alarm with activated buzzer (chap. 8.1), by the buzzer sounding), proceed as

follows:

Disconnect the unit from the power supply.

Have an expert examine and rectify the cause of the fault.

Restart the unit (see chapter 5).


Adjustment:
To check the response temperature of the safety thermostat class 3.1, turn on the chamber and set the

desired set point on the temperature controller.
The sections of the scale from 1 to 10 correspond to the temperature range from 0 °C / 32 °F to 120 °C /

248 °F (BD) or from 63 °C / 145.4 °F to 350 °C / 662 °F (ED or FD, option) and serve as a setting aid.

1. Turn the control knob (7) of the safety device with a coin to its

end-stop (chamber protection).

2. When the set point is reached, turn back the control knob (7)

until its trip point (turn it counter-clockwise)

3. The trip point is identifiable by the red alarm lamp (7a) lighting

up.
With the option audible alarm and the buzzer activated (chap.

8.1), the buzzer sounds as an additional signal. You can turn it

off with switch (10) resp. (9).

4. The optimum setting of the safety device is obtained by turning

the control knob clockwise by approx. one scale division, which

leads to extinguish the red alarm lamp (7a).

Check the setting regularly and adjust it following any changes of the set-point.

Function check:
Check the temperature safety device class 3.1 at appropriate intervals for its functionality. It is

recommended that the authorized operating personnel should perform such a check, e.g., before starting

a longer work procedure.