BD / ED / FD (E2) 02/2015

page 51/63

Unit size

FD 23

FD 53

FD 115

FD 240

Temperature data
Temperature range, 5 °C above ambient

up to

°C / °F

300 / 572 300 / 572 300 / 572 300 / 572

Temperature fluctuation at 150 °C / 302 °F

± K

0.3

0.3

0.3

0.3

Temperature

uniformity

(variation 1)

at 70 °C / 158 °F

± K

0.8

0.8

0.7

0.8

at 150 °C / 302 °F

± K

2.2

2

1.8

2

at 300 °C / 572 °F

± K

4.3

3.7

3.9

4.3

Heating up time 2)

to 70 °C / 158 °F

Min

7

7

7

11

to 150 °C / 302 °F

Min

22

24

28

24

to 250 °C / 482 °F

Min

45

60

49

50

Recovery time

after door was

open for 30 sec

2)

at 70 °C / 158 °F

Min

2

2

2

2

at 150 °C / 302 °F

Min

4

5

5

6

at 300 °C / 572 °F

Min

9

9

12

13

Ventilation data

Air change

at 70 °C / 158 °F

x/h

59

59

29

19

at 150 °C / 302 °F

x/h

64

64

32

20

at 300 °C / 572 °F

x/h

53

53

26

18

Electrical data
IP system of protection acc. to EN 60529

IP

20

20

20

20

Nominal voltage (±10%) 50/60 Hz

V

230 1N~ 230 1N~ 230 1N~ 230 1N~

Nominal power

kW

0.8

1.20

1.60

2.70

Power plug

shock proof plug

Unit fuse

5x20mm / 230V / time-lag M

Amp

10

external

10

external

10

external

16

external

Energy

consumption

at 70 °C / 158 °F

Wh/h

145

172

230

370

at 150 °C / 302 °F

Wh/h

300

429

544

850

at 300 °C / 572 °F

Wh/h

720

951

1100

1400

Installation category acc. to IEC 61010-1

II

II

II

II

Pollution degree acc. to IEC 61010-1

2

2

2

2

Electrical connection data FD-UL acc. to cUL standard (for USA and Canada):

Unit size

FD 23-UL FD 53-UL FD 115-UL FD 240-UL

Electrical data
Nominal voltage (±10%) 60 Hz

V

115 1N~ 115 1N~ 115 1N~ 208 3N~

Nominal power

kW

0.80

1.20

1.60

2.70

Nominal current

Amp

7.0

10.5

14.0

11.2

Power plug

NEMA

5-15P

5-20P

5-20P

L21-20P

Unit fuse

6,3 x 32 mm / 250V / super time-lag TT

Amp

12,5

external

16

external

16

external

3 x 16

internal

Legend: 1) without outer glass door 2) up to 98 % of the set value
All technical data is specified for unloaded units with standard equipment at an ambient temperature of

+25 °C / 77 °F and a power supply voltage fluctuation of ±10. The temperature data is determined in

accordance to BINDER factory standard following DIN 12880, observing the recommended wall

clearances of 10 % of the height, width and depth of the inner chamber. Technical data refer to 100% fan

speed.
All indications are average values, typical for units produced in series. We reserve the right to

change technical specifications at any time.

If the cabinet is fully loaded, the specified heating up times may vary according to the load.