BD / ED / FD (E2) 02/2015

page 45/63

11. Troubleshooting

Fault description

Possible cause

Required measures

Temperature

Set-point temperature is not

reached after specified time.

Unit door not properly closed. Completely close unit door.
Door gasket defective.

Replace door gasket,

Controller not adjusted, or

adjustment interval exceeded. Calibrate and adjust controller.
Wrong voltage.

Check power supply for voltage of

115V or 230V.

FD: The fan doesn’t turn or turns

too slowly.

Fan defective.

Contact BINDER service.

Chamber heating permanently,

set-point not held.

Controller defective.

Contact BINDER service.

Pt 100 sensor defective.
Semiconductor relay defective
Controller not adjusted, or

adjustment interval exceeded. Calibrate and adjust controller.

Chamber doesn’t heat up.

Red heating control light in the

display is lit.

Heating element defective.

Contact BINDER service.

Semiconductor relay defective.

Chamber doesn’t heat up.

Red heating control light in the

display is not lit.

Controller display working.

Timer run off.

Program the timer or change to time

function Continuous operation

(chap. 6.3)

Semiconductor relay defective.

Contact BINDER service.

Controller defective.

Unit without function, only the

green “stand-by” LED is lit

Unit in stand-by mode

Press down the ON/OFF button (5)

until the display lights up.

BD, option ED,FD: Temperature

inside the chamber too high, Red

alarm pilot lamp of safety device

(7a) is lit..

Safety device class 3.1 has

responded.

Check the settings of the

temperature set-point and of the

safety device class 3.1 (chap. 7.2).

ED, FD: Unit without function.

Red alarm pilot lamp of safety

device (7a) is lit.

Safety device class 2 has

turned off the chamber.

Let cool down the chamber and

press down RESET button. Check

the settings of the temperature set-

point and of the safety device class

2 (chap. 7.1). If appropriate, select

suitable limit value.

Safety device class 2

defective.

Contact BINDER service.

Unit without any function.

No power supply.

Check connection to power supply.

Unit fuse has responded.

Check unit fuse and replace it if

appropriate. If it responds again,

contact BINDER service.

Controller defective.

Contact BINDER service.

Deviations from the indicated

heating-up times.

Chamber fully loaded.

Load the chamber less or consider

longer heating-up times.

Controller
Message „1999” in the controller

display

Sensor rupture between

sensor and controller.

Contact BINDER service.

The controller returns to Normal

Display from any level.

No button was hit for more

than approx. 30 sec.

Repeat entries, enter the values

rapidly.

Only qualified service personnel authorized by BINDER must perform repair.

Repaired units must comply with the BINDER quality standards.