5 general notes – BINDER BD 23 User Manual

Page 31

Advertising
background image

BD / ED / FD (E2) 02/2015

page 31/63

1. Press down

button for approx. 5 seconds.

The display alternately shows “unit” and the temperature unit:

2. Press again button

.

The display alternately shows “rASd” and the set-point gradient:

3. Press again button

.

The display alternately shows “

Adr” and the actual setting of the unit address:

4. Set the required address with buttons

.

You can enter address values between 1 and 30.

5. The set value is automatically adopted after 2 seconds.


6.5 General notes

Approx. 30 sec. after the last entry the controller returns to normal display (actual value

display).

The functions set-point entry (chap. 6.1), time functions (chap. 6.3), and calling up the user

menu (chap. 6.4) can only be selected from normal display (actual value display).

When selecting the functions set-point entry and time functions, and when selecting the user

menu functions, the respective button

or

must be pressed down for a about 1 sec.

Shorter pressing will be ignored by the controller.

After a power failure, the timer returns to the previous status. A remaining time, if any, will

continue running down.

Adjust the safety device following any changes of the set-point (chap. 7).