To Enter a New Setup and Calibration Number

Short Form Calibration

The Short Form Setup & Calibration procedure allows you to change the “SETUP” and “CAL”
numbers of the indicator. You may want to perform this procedure if the indicator is being
connected to different load cells or if the scale is inaccurate and you have some accurate known
weights to use for calibrating the scale.
Before continuing, first write down the current SETUP and CAL numbers of your EZ indicator.
These numbers are displayed during the Self Test.

To run the self test with the indicator already ON:

Press [NET/GROSS] and then press [ON/RECORD] to start the self-test.
Press [ON/RECORD] to pause the self-test to view the SETUP and CAL numbers.
Press [ON/RECORD] again to allow the self-test to complete normally.

.SETUP # _______________ CAL # ________________

CAUTION: Do not attempt to calibrate the scale if the indicator is not reading stable weights.

The calibration procedure will not fix instability, inconsistencies, or flashing
"RANGE" messages.


Hold [ZERO/RECHECK] and hold [ON/RECORD] for 3 seconds to enter the short form calibration.
The display will flash “SETUP” and then display the 6-digit setup number with the right digit flashing.
To modify the setup number:

1. Press [Net/Gross] several times to increment the digit to the proper value.
2. Press [TARE] to advance the blinking digit to the left.
3. Repeat steps 1 and 2 for each digit as required.
4. Press [ON/RECORD] to enter the new setup number and display the calibration number.
5. Repeat steps 1 and 2 to modify the calibration number.
6. Press [ON/RECORD] to enter the new calibration number and the display will go back to

normal.

7. Verify the accuracy of the scale and the procedure is complete!