Troubleshooting Guide

Disconnect the first load cell. Connect the

other load cell to the indicator (use the

same connector). Stand over this load

cell. Does the indicator display a reading

approximately twice your weight?

The other load cell

is defective. Check

it's cable for cuts or

flattened areas and

send it back for repair.

The indicator is defective. Verify this

by substituting a good working indicator.

Check the platform for

debris and mud

interference. Check

mounts for binding and

obstructions.

Connect the load cell to the

other connector on the indicator.

Stand on the load cell, Does it

display approximately twice your

weight?

This load cell is defective.

Check it's cable for cuts or

flattened areas or send it

back for repair. Double check

the indicator using a known

good load cell.

This load cell is OK. Disconnect the

first load cell and connect the other

load cell to the same connector.

Now stand over the second load cell.

Does it weigh twice your weight?

Disconnect one load cell cable from the indicator. Press

the indicator "TARE" button. (The indicator display

should go to zero) Stand on the load cell. Does the

indicator display a reading about twice your weight?

Connect both load cells (one at a

time) to the other connector on the

indicator. Do both load cells weigh

you about twice your weight?

One of the load cells is bad. A

good load cell will weigh you at

approximately twice your weight with

only one load cell connected.

Do both load cells weigh you

about the same but not within

20% of twice your weight?

Does the indicator turn on?

Does the scale weigh you

close to your weight?

Does the indicator respond when

you step on the scale?

Check for 12VDC at the AC/DC converter power cord. If you do

not have a multimeter, use a car battery to power the scale. If

the indicator still does not come on, send it in for repair.

Check for debris

around and under the

platform and load

cells. Check for

warped or damaged
mounts on platform.

Check for debris under

or around the scale
mounts or platform.

Check the load cell
cables for cuts and

flattened areas.

The other load cell is

defective. Check it's cable

for cuts or flattened areas or

send it backl for repair.

Check the defective load cell

by connecting it to the

indicator and verify that it

does not weigh.

Disconnect the first load cell

and connect the other load
cell to the same connector

on the indicator. Does the
indicator display a reading

about twice your weight.

Disconnect one of the load cells and stand over

the connected load cell. Does the indicator

display a reading about twice your weight?

Perform the paper clip test. Take

two paper clips and carefully insert

one into pin 2 and the other in pin 3
of either load cell connector jack on
the indicaor. Hold one paper clip in

each hand (If they touch it will not

damage the indicator.) Does the

reading on the indicator jump

around?

Both load cells are defective.

The indicator is probably OK.

Check the cables for damage and

send both load cells in for repair.

The indicator is defective. Send

it in for repair. Note: Load cells

may also be defective. Check the

load cells by attaching to a good

working indicator.

The other load cell is defective.

Check it's cable for cuts or flattened

areas and send it back for repair.