1 ca-6 technical specifications – Electro-Chemical Devices (ECD) CA6 Hardness Analyzer User Manual

Page 13

Advertising
background image

1.0.1 CA-6 Technical Specifications

Analysis:

Colorimetric parameters

Method:

Photometric differential absorbance

Measuring range:

0.00 – 1.00 ppm configured 000 – 1000 ppb

Response time:

6 minute 10 second cycle plus wait time

Repeatability:

+/- 2% on absorbance value with turbidity < 80 NTU

Drift:

+/- 2% per month on the absorbance measurement

Power supply:

110-220Vac, 50-60 Hz 80 VA

Mounting:

Wall mounting or with optional bench support

Operating temperature:

5°C to 50°C

Cabinet: Stainless steel, epoxy powder coated

Dimensions:

15”W x 24”H x 8.25”D (380mm W x 600mm H x 210mm D)

Weight:

Approx. 38 lbs (17 kg)

Reagent consumption:

1 liter each reagent in 30 days of continuous operation with no wait time
between cycles

Analog output:

4-20 mA

Alarms:

4 configurable relays

Sample

Inlet sample pressure:

Atmospheric

Outlet sample pressure:

Atmospheric, waste tubing

⅜” O.D. x ¼” I.D.

Sample flow for the fast loop
reservoir:

100-500 ml / min

Connections:

To the fast loop reservoir with flexible tubing ¼”O.D. x

⅛” I.D.

2