Fill a 1 liter volumetric flask to 2/3rds capacity with DI water. Add 200 ml of ethanolamine while stirring
the solution. Fill the volumetric flask to the gauge line (1 liter). Transfer to the reagent 2 container
provided with the analyzer.

11.4.3 Standard Solution
Standard Solution, 100 ppm as CACO

3

Part Number# 2010039-1

Calibration solutions can be made by diluting a 100 mg/l Hardness solution with an appropriate volume
of deionized water, 1.0ml of 100 mg/l (ppm) in 1 liter of water corresponds to a 100 ppb solution of
CACO

3

.

Preparation Standard solution 0.1 g/l = 100 mg/l

CACO

3

Dissolve 0.146 g CaCl

2

.2H

2

O in 1 liter of DI water

By diluting an appropriate volume of the standard solution with DI water, various calibration solutions
with a known hardness concentration may be prepared.

1.0 ml Stock into 1.0 liter DI water = 100 ppb Hardness standard solution

10.0 ml Stock into 1.0 liter DI water = 1000 ppb Hardness standard solution