11.2 Extra Cycle

The EXTRA CYCLE is programmed as a BLANK but can be programmed as a CALIBRATION or VALIDATION
Cycle. The EXTRA CYCLE is not activated when shipped to activate the EXTRA CYCLE go to Program >
Settings > Cycles Ratio, see section 8.6.3. The programmed sequence for the Analysis Cycle is detailed in
the following table. This program assumes that the Sample is connected to the Left side of the selection
valve (Sample Port #1) and the Calibration solution or Deionized water is connected to the Right side of
the selection valve (Sample Port #2).

Step

Seconds

Function

1

5

DRAIN

2

10

RINSE #2

3

16

SAMPLE #2

4

5

DRAIN

5

16

SAMPLE #2

6

5

DRAIN

7

16

SAMPLE #2

8

5

DRAIN

9

10

SAMPLE #2

10

10

WAIT

11

2

REFERENCE

12

12

ADD REA #1

13

12

ADD REA #2

14

150

MIX

15

10

WAIT

16

1

ABSORBANCE

17

2

BLANK (Validation or Calibration)

18

0

WAIT

19

0

WAIT

20

5

DRAIN

21

16

SAMPLE #1

22

5

DRAIN

23

16

SAMPLE #1

24

5

DRAIN

25

13

SAMPLE #1

26

0

DRAIN

27

0

SAMPLE #1

28

0

WAIT

29

0

WAIT

30

10

WAIT (User adjustable time in seconds)

Step 30 is adjustable by the end user: go to PROGRAM >SETTINGS >Cycle Wait