Electro-Chemical Devices (ECD) CA6 Hardness Analyzer User Manual

Page 25

Advertising
background image

Add rea #3
Add rea #3 turns on the micro peristaltic pump for the addition of reagent

#3. The mixing pump is circulating the sample as the reagent is being

added. The Add Rea #3 time is set in seconds.

Add rea #4
Add rea #4 turns on the micro peristaltic pump for the addition of

reagent #4. The mixing pump is circulating the sample as the reagent is

being added. The Add Rea #4 time is set in seconds

.

Wait

The waiting time is set in seconds. Wait puts the CA6 in Stand By mode.
Mix

The mixing time is set in seconds.
Reference

The reference measurement is the first point for the calculation of the absorbance, I

0

. Reference sets the

time, in seconds, of when to take the base intensity measurement.
Absorbance

Absorbance sets the time, in seconds, of when to take the colored intensity measurement, the Reading,
I

1

, and calculate the absorbance, I

0

/ I

1

. Typically 1 – 2 seconds

Calculation

Calculation sets the time in seconds to convert the absorbance reading into a concentration reading and
sends the calculated value to the main display. Calculation uses the absorbance reading and the
calibration factor. Typically 1 – 2 seconds

14