8.4 RUN MENU

The RUN Menu provides access to the Start /Stop
functions of the analyzer, Start On-Line, Cycles to Run
and Emergency Stop.

8.4.1 START ON-LINE?
By pressing this button the analyzer will start a
continuous cyclic analysis based on the steps set in the
Analysis Cycle of the PROGRAM menu. ONLINE is now
displayed on the button. By pressing the button again,
the analyzer will finish the cycle in progress and then wait in STAND-BY mode.

8.4.2 CYCLE Ch1 / CYCLE Ch2 / Cycle Extra
This command starts the analyzer, performing one cycle of analysis on channel 1 or channel 2 or extra
auto-function. At the end of the cycle, the analyzer will turn in stand-by conditions waiting for a new
user’s command.

8.4.3 EMERGENCY STOP
This command will immediately stop the analyzer at the current step of the analysis. The analyzer will go
into MANUAL STOP condition.

MANUAL STOP

in red colored letters will appear on the main screen and

the FAULT contact will be activated. To restart the analyzer after an emergency stop is necessary to
reset the fault. Press RUN then RESET EM. STOP yellow colored letters and the analyzer will switch to
STAND-BY mode.

35