10.0 ANALYZER SHUT DOWN

If the CA-6 Analyzer will be out of service for a period of two weeks or greater, proceed as follows:

1.

Empty all reagent containers.

2.

Rinse and refill all reagent containers with distilled water.

3.

Prime all of the pumps with DI water. (login with 1

st

level password, press DISPLAY,

MANUAL STEP, ADD REAG 1,2,3,4 for 40 seconds each)

4.

Disconnect the sample feed line and fill the fast-loop reservoir (if present) with distilled
water.

5.

With the sample inlet tubing attached to a container of distilled water, run the analyzer
for at least 2 cycles.

6.

Empty the water from all lines.

7.

Put the analyzer in stand-by condition.

8.

Turn OFF the power to the analyzer and disconnect the plug from the wall socket.