1 analysis cycle – Electro-Chemical Devices (ECD) CA6 Hardness Analyzer User Manual

Page 76

Advertising
background image

11.1 Analysis Cycle

The programmed sequence for the Analysis Cycle is detailed in the following table. This program
assumes that the sample is connected to the Left side of the selection valve (Sample Port #1) and
Deionized water or Calibration solution is connected to the Right side of the selection valve (Sample Port
#2).

Step 30 is user configurable to allow up to 15 minutes (900 seconds) between Analysis Cycles. To enter a
time other than 30 seconds go to PROGRAM > SETTINGS > Cycles Wait and enter the new time in
seconds.

Step

Seconds

Function

1

5

DRAIN

2

10

RINSE #1

3

16

SAMPLE #1

4

5

DRAIN

5

16

SAMPLE #1

6

5

DRAIN

7

16

SAMPLE #1

8

5

DRAIN

9

10

SAMPLE #1

10

10

WAIT

11

2

REFERENCE

12

12

ADD REA #1

13

12

ADD REA #2

14

150

MIX

15

10

WAIT

16

1

ABSORBANCE

17

2

CALCULATION CH1

18

0

WAIT

19

2

SAVE DTLOG

20

5

DRAIN

21

16

SAMPLE #1

22

5

DRAIN

23

16

SAMPLE #1

24

5

DRAIN

25

13

SAMPLE #1

26

0

DRAIN

27

0

SAMPLE #1

28

0

WAIT

29

0

WAIT

30

10

WAIT (User adjustable time in seconds)

Step 30 is adjustable by the end user: go to PROGRAM >SETTINGS >Cycle Wait

65

Advertising