7 service menu – Electro-Chemical Devices (ECD) CA6 Hardness Analyzer User Manual

Page 50

Advertising
background image

8.7 SERVICE MENU

This menu is protected with a 2

nd

level

password and contains settings critical to
the performance of the analyzer. These
values should only be modified by
properly trained personnel.
Service Window #1
Unit
: Choose the unit of measurement:
ppm or ppb.

Sensor av.: Set the number of cycles averaged for the sensor reading.

Cal. Factor MIN: Minimum allowable calibration factor.

Cal. Factor MAX: Maximum allowable calibration factor.

Blank: Enter the blank value manually (see section 7.1). Use the value in Section 11.3 as a default
setting.

Method: Choose the specific measured parameter (N/NH3, PO4, N/NO2, SiO2...).

F. Scale1/2: Set the full scale value for the 4-20 mA output (0 to 200.0 ppb = 4-20 mA, above)

Reference Min: This sets the minimum reference value allowable. Values below this value cause a
FAULT. This failure is typically caused by a coating on the cell blocking the light transmittance, clean the
cell.

Bottle ml: Enter the volume of the reagent(s) bottle(s).

Cycle sec.: Displays the duration of the programmed analysis cycle.

Ext. Input: Allows a selected function to be executed from an external input: extra cycle, on line, print or
none.

Relay #1 - #4: Configure the functions of the relays.

Cycle Command (relay activated as programmed in the steps in the PROGRAM
menu)

Fault Alarm (relay activated in case of fault alarm)

Loss of Sample (relay activated in case of loss of sample alarm)

39

Advertising