3 accessories and spare parts – Electro-Chemical Devices (ECD) CA6 Hardness Analyzer User Manual

Page 72

Advertising
background image

9. Position the tubing and fittings deep into the pump

head body.

10. Using Manual Mode, activate the peristaltic pump for 1

second and insert one roller at a time in the free
position (where the roller is not in contact with the
tubing). Repeat for the 3 rollers.

11. Replace the Plexiglas pump cover.

Attach the cover

with the 3 screws TSPEI M3x8 using an Allen Key.

12. Using Manual mode activate the peristaltic pump for

180 seconds (3 minutes) to settle the tubing into
position.

13. Reconnect the outlet tubing to the peristaltic pump fitting.
14. Reconnect the reagent intake tubing to the peristaltic pump fitting.
15. Using Manual mode activate the peristaltic pump for 45 seconds to prime the pump.
16. Repeat the above steps for each additional pump for which it is necessary.

Maintenance frequency for tubing replacement is 70 working hours. To calculate the maintenance
frequency for a pump, simply determine the number of analysis cycles per day and the reagent's dose
time for each analysis cycle.

Example: 4 analysis cycles per hour (i.e. 96 analysis cycles per day), Dose time 1 5 seconds for each
analysis cycle. Working hours per day = (96 cycles/day)(15 sec) / 3600 sec/hour = 0.4 hours / day

Maintenance frequency = 70 hours / 0.4 hours / day = 175 days

9.3 Accessories and Spare Parts
Model CA6 Analyzer: Spare Parts

Part#

Description

2000005

*Recommended Spare Part

Yearly Maintenance Kit, Norprene size #16 tubing (4 each), Pinch Valve Tubing (6 each),

Colorimetric Cell O-Rings (2 each) Tubing Replacement Tool (Sample Pump)

2010036-1 Hardness Analyzer, 2 Reagent Kit, 2 x 500 ml bottles

2010037-1 Hardness Reagent #1 , 500 ml bottle

2010038-1 Hardness Reagent #2 , 500 ml bottle

2010039-1 Calibration Solution, Hardness, 100 ppm, 2 liter bottle

9380012

Mixing Pump - PP EPDM

9380013

Mixing Pump - PP Teflon

9370044

Pinch Valve

9370045

3-Way Valve

3000004

Support for 3-Way Valve

8700001

Heater Element (Thermostatic Block)

3000001

Quartz Reaction Cell 16 mm *Recommended Spare Part

61