Installation inspection of the reference machine – Electrolux FOM 71 CLS User Manual

Page 12

Advertising
background image

3:2

Installation inspection of the reference machine

5330

1

Drain connection

• Connect a 75 mm (2 1/2") flexible rubber hose

to the machine’s drain pipe, ensuring a down-
ward flow from the machine. Avoid sharp
bends, which may prevent proper drainage.

• There should be an air-gap of 50 mm between

the drain hose and the drain.

• Make sure that the drain hose is sufficiently

flexible so that it does not cause problems
related to the scale system.

• Pipes made from PVC or similar materials

may not be connected to the machine due to
the load cells.

• The drainage hose should be located over a

floor drain, drainage channel or similar so that
the distance between the outlet and the drain
is at least 25 mm (1").

• Check that the amount of supply water at the

water connection to the reference machine is
sufficient and compliant with the specifica-
tion in the standard (15 ± 2 litre) and that the
amount can be maintained at all times. The
water pressure at the intake of the machine
may never fall below 100kPa.

• Check that the power supply is compliant with

the voltage specified on the reference machi-
ne data label and within the tolerances of the
standard (± 2%) .

• Make sure that the power supply can maintain

the required voltage level whilst the reference
machine is heating.