Fom 71 mp-lab empty run test data, Empty run test data – Electrolux FOM 71 CLS User Manual

Page 83

Advertising
background image

24:1

FOM 71 MP-Lab

Empty run test data

During the CLS MP-Lab ring test, data was obtained from a number of laboratories.
This data is useful when checking that the reference machine in the laboratory is
performing as expected.
Please note: This data is obtained from runs with an unloaded machine. Therefore,
the energy and process times obtained here cannot be compared with the same
parameters applied during a test series carried out with a loaded machine. However,
empty test runs are not affected by load differentials or detergent batches and will
only produce results that are specific to the machine.
Table 1 below shows the estimated/defined values for Cotton 60°C, as well as av-
erage values from the RRT 2008 ring test.
The RRT 2008 average values were obtained from the RRT 008 IEC/CENELEC/CE-
CED ring test. These values are based on empty run tests undertaken at 15 different
laboratories.
The definitions used in Table 1 differ from those given in IEC 60456 and below are
applicable to Table1.

Definitions
Total program time:

Time from the start of first filling to the start
of the retardation phase at the end extraction
sequence.

Heating time:

The time until the heating relays have been
switched on during the different heating peri-
ods.

Total main wash time:

The time from start of first filling until the start
of first rinse water supply.

Filling time of water:

The first filling time (26 litre). From closure of
door lock relay to end of filling.

Cold water inlet temperature:

The temperature of the supplied cold water.

Temp at end of heating in

enzyme step:

The temperature at the time the heating relay is

switched off after reaching the 40ºC enzyme

step.

Temp at the end of the

enzyme step:

The temperature in the bath just

prior to the heating up to 60ºC.