Calibration of level control system

5

• Close the door.
• Go down to ”Calibration of level sensor”.
• Press ”Select”.

• Press ”Open” to fill the machine with water until the level reaches the botton

of the inner drum where the ruler is placed.

• Press ”OK”.
• Enter the value read on the ruler in RU units (The first setting is 0=level in

drum.

• Press ”OK”.
• Press ”Open” to fill more water.
• Press ”OK”.
• Enter the value read on the ruler in RU units. (If you enter an incorrect value,

press ”CLR”. Enter new value.

• Press ”OK”.
• Enter another three values, making five in total.
• Once all values have been entered, press ”Save”.

LEVEL CALIBRATION

FILL WATER TO

DESIRED LEVEL

LEVEL:

0 SCU

OPEN

OK

5