Electrolux FOM 71 CLS User Manual

Page 71

Advertising
background image

20:1

Flow meter: maintenance

FOM 71 MP-Lab

Flow meter: maintenance

(All references in this document refers to ELS document 438 9041-07 Instal-
lation of flow meter control FOM 71 LAB. See Appendix 1)
• Occasionally check the amount of water coming from the flow meter. If

the amount is not correct, recalibrate in accordance with the procedures
set out in Chapter 16.

• The impeller in the flow meter is sensitive to contaminants in the water

and after long inoperative periods, the impeller may stop functioning.

• If the impeller is not working properly, or if the connection between im-

peller or control unit is broken, the water intake will stop at 8 litre.

• In this case, check the electrical connection between the impeller and

control unit or or ascertain where the impeller may be stuck open and
clean the impeller in accordance with the following instructions:
o Remove the electrical housing (Unscrew the side mounted screw and

twist and lift off the housing.)

o Unscrew the 4 screws on the impeller housing and remove the impel-

ler plate with impeller.

o Carefully turn the impeller and rotate (blow on) the impeller until it

rotates free.

o Refit the impeller and electrical housing.

Advertising
This manual is related to the following products: