Constant current output, Constant current output -6 – RIGOL DP832A User Manual

Page 60

Advertising
background image

RIGOL

Chapter 2 Front Panel Operations

2-6

DP800A User’s Guide

Constant Current Output

In constant current output mode, the output current equals the current setting value

and the output voltage is determined by the load. This section introduces the

operation method in constant current output mode.

Operation Method:

1. Connect the channel output terminals

Connect the load with the channel output terminals of the corresponding

channel at the front panel by referring to Connect the channel output

terminals in “Constant Voltage Output”.

2. Turn on the power switch to start the instrument.

3. Select the channel

Select the proper output channel according to the desired output current. Press

the corresponding channel selection key; at this point, this channel, its channel

number and output state are high-lighted in the center of the screen.

4. Set the voltage

Press Voltage to set the desired voltage according to Set the voltage in

“Constant Voltage Output”.

5. Set the current

Press Current to set the desired current according to Set the current in

“Constant Voltage Output”.

6. Set the overvoltage protection

Press OVP to set the proper overvoltage protection value (for the setting

method, refer to Set the voltage in “Constant Voltage Output”). Then,

enable the overvoltage protection function (you can enable or disable the OVP

function by pressing OVP) and the output will be turned off automatically when

the actual output voltage is greater than the overvoltage protection value.

7. Turn on the output

Turn on the output of the corresponding channel and the actual output voltage,

output current, output power as well as the output mode (CC) are high-lighted

in the user interface.

Warning

To avoid electric shock, please turn on the output switch after the

output terminals are correctly connected.