RIGOL DP832A User Manual

Page 95

Advertising
background image

Chapter 2 Front Panel Operations

RIGOL

DP800A User’s Guide

2-41

3. Trigger Condition

Press Condition to open the trigger condition setting interface and users can

set the trigger condition of trigger output.

 Output Trig: the instrument triggers when the output of the control source

is turned on or off.

Press Outp Trig and use the up/down direction key to select “OutpClose”

or “OutpOpen”.

 Voltage Trig: the instrument triggers when the output voltage of the control

source meets the trigger condition set.

Press Volt Trig and use the up/down direction key to select “>”, “<” or “=”;

use the numeric keyboard or the left/right direction key and knob to set the

voltage value.

 Current Trig: the instrument triggers when the output current of the control

source meets the trigger condition set.

Press Curr Trig and use the up/down direction key to select “>”, “<” or “=”;

use the numeric keyboard or the left/right direction key and knob to set the

current value.

 Power Trig: the instrument triggers when the output power of the control

source meets the trigger condition set.

Press PowerTrig and use the up/down direction key to select “>”, “<” or

“=”; use the numeric keyboard or the left/right direction key and knob to

set the power value.

 Auto Trig: when the trigger output is enabled, the instrument triggers

automatically.

4. Output Signal

Press Signal to set the signal type of the trigger output to “Level” or “Square”.
 Level: the output high level is from 2.6V to 3.5V and the output low level is

from 0V to 0.4V.

 Square: press Period to set the square period and the range is from 100μs

to 2.5s; press Duty to set the duty cycle of the square waveform and the
range is from 10% to 90%.

5. Polarity

Press Polarity to set the polarity of the trigger output signal to “Positive” or

“Negative”.
 Positive: output the signal selected in Signal when the trigger condition is

met.

 Negative: turn the signal selected in Signal upside down and then output

the signal when the trigger condition is met.

6. Enable

Press Enable to enable or disable the trigger output function of the

corresponding data line.

Advertising
This manual is related to the following products: